Microscopy Resolution and Image Formation
Two fluorescent emitters can be physically separate but still appear as one blurred spot. Change their separation, wavelength, numerical aperture, and photon number to see how the point-spread function and photon noise determine what the microscope can actually resolve.
Rayleigh criterion 0.61λ/NA
264 nm
Gaussian PSF FWHM
221 nm
Separation / Rayleigh limit
0.95
Interpretation
Strong overlap
True Object Plane
What the Camera Records
Intensity Profile Through Both Emitters
Main ideas
A point emitter is recorded as a finite-width point-spread function rather than a mathematical point.
Shorter wavelength and larger numerical aperture narrow the PSF.
Photon number changes image noise and confidence, but collecting more photons does not make the diffraction-limited PSF itself narrower.